Uji Kesesuaian Kualitas Berkas Sinar-X Menggunakan Half Value Layer (HVL) Filter Aluminium Pada Pesawat Sinar-X Stationary di Akademi Teknik Radiodiagnostik dan Radioterapi Bali

  • Ni Putu Pande Anita Yulianti Udayana
  • Gusti Ngurah Sutapa Udayana University
  • Kadek Yuda Astina Akademi Teknik Radiodiagnostik dan Radioterapi Bali, Jl. Tukad Batanghari VII No.21, Dauh Puri Klod, Denpasar Selatan, Bali, Indonesia 80225

Abstract

A study was conducted to determine the quality of the X-ray beam using a half value layer (HVL) aluminum filter on a stationary X-ray device at the Akademi Teknik Radiodiagnostik dan Radiotherapi (ATRO) Bali. This study aims to determine the quality of the X-ray beam on a stationary X-ray device at ATRO Bali by looking at the magnitude of the HVL value produced. The tools used are aluminum filters with thicknes of 0.5, 1.0, and 2.0 mm which are varied for each current used, a constant tube voltage of 80 kV, tube currents of 100, 125, 160, and 200 mA, exposure time of 0.1 s and an X-ray multimeter. The value of HVL was determined by linear regression test using SPSS. On each tube current used, the average HVL value was 2.77 mmAl. The t-test indicated that the HVL value is not in accordance with the KMK standard number 1250 in 2009 which is in the range of ? 2.3 mmAl to < 2.5 mmAl.


Abstrak


Telah dilakukan penelitian untuk mengetahui kualitas berkas sinar-X menggunakan half value layer (HVL) filter aluminium pada pesawat sinar-X stationary di Akademi Teknik Radiodiagnostik dan Radioterapi (ATRO) Bali. Penelitian ini bertujuan untuk mengetahui kualitas berkas sinar-X pada pesawat sinar-X stationary di ATRO Bali dengan melihat besarnya nilai HVL yang dihasilkan. Alat bantu yang digunakan berupa filter aluminium dengan tebal 0,5, 1,0, dan 2,0 mm, yang divariasi pada setiap besarnya arus yang digunakan, tegangan tabung konstan sebesar 80 kV, arus tabung 100, 125, 160, dan 200 mA, waktu eksposi sebesar 0,1 detik dan multimeter sinar-X. Untuk mengetahui besar nilai HVL digunakan uji regresi linier menggunakan SPSS. Pada masing-masing arus tabung yang digunakan diperoleh nilai HVL rata-rata 2,77 mmAl. Hasil uji-t menunjukkan bahwa nilai HVL tersebut kurang sesuai dengan standar KMK nomor 1250 tahun 2009 yaitu dalam rentang ? 2,3 mmAl sampai < 2,5 mmAl.

Downloads

Download data is not yet available.

References

[1] D. H. Yunitasari, E. Setiawati, dan C. Anam, Evaluasi Metode Penentuan Half Value Layer (HVL) Menggunakan Multi Purpose Detector (MPD) Barracuda Pada Pesawat Sinar-X Mobile, Jurnal Fisika, vol. 3, no. 2, 2014, pp. 113-118.
[2] Keputusan Menteri Kesehatan, Pedoman Kendali Mutu (Quality Control) Peralatan Radiodiagnostik, Menteri Kesehatan Republik Indonesia, 2009, pp. 19-21.
[3] Badan Pengawas Tenaga Nuklir, Uji Kesesuaian Pesawat Sinar-X Radiologi Diagnostik dan Internasional, Badan Pengawasan Tenaga Nuklir Republik Indonesia, 2018, pp. 62.
[4] M. Akhadi, Dasar-Dasar Proteksi Radiasi, Jakarta, 2000.
[5] K. Miniati, G. N. Sutapa, dan B. Sudarsana, Uji Kelayakan Pesawat Sinar-X Terhadap Proyeksi Pa (Postero-Anterior) dan Lat (Lateral) Pada Teknik Pemeriksaan Foto Thorax, Jurnal Fisika, vol. 18, no. 1, 2017, pp. 27-31.
[6] L. G. P. Satwika, N. N. Ratini, dan M. Iffah, 2021, Pengaruh Variasi Tegangan Tabung Sinar-X Terhadap Signal to Noise Ratio (SNR) Dengan Penerapan Anode Heel Effect Menggunakan Stepwedge, Jurnal Fisika, vol. 22, no. 1, 2021, pp. 20-28.
[7] F. Suyatno, L. Yuniarsari, dan B. Syawaludin, Rancangan Sistem Instrumentasi Pembangkit Sinar-X Pada Pesawat Rontgen, Jurnal Batan, vol. 5, no. 2, 2009, pp. 151-165.
[8] N. Mukminah, R. Iswadi, dan Ihsan, Analisis Linearitas Keluaran Radiasi Pada X-Ray Mobile Dengan Menggunakan Piranha, Skripsi, Program Studi Fisika, Fakultas Sains dan Teknologi Universitas Islam Negeri Alauddin Makasar, 2014.
[9] S. Litasova, E. Hidayanto, dan M. Azam, Pengaruh Ketebalan Dan Kombinasi Jenis Filter Terhadap Nilai Entrance Skin Exposure (ESE) Menggunakan Faktor Eksposi Pemeriksaan Kepala, Jurnal Fisika Undip, vol. 7, no. 2, 2018, pp. 67-75.
[10] H. Dabukke, Pengujian Iluminasi, Kolimasi, Ketegaklurusan dan Kualitas Berkas Pesawat Sinar-X Radiografi Umum Dengan Radiografi Mobile, Tugas Akhir, Program Studi Fisika, Fakultas Matematika dan Ilmu Pengetahuan Alam Universitas Sumatera Utara, 2018.
[11] A. Mutmainna, D. S. Astuty, S. Dewang, dan Mulyadin, Uji Kesesuaian Standar Nilai HVL Filter Aluminium Pada Pesawat Sinar-X Mammografi Studi Kasus Di Ruang Instalasi Radiologi RS. Siloam Makasar, Jurnal Fisika, vol. 23, no. 1, 2020, pp. 17-25.
[12] D. K. Lestari, Studi Penentuan Kualitas Berkas Radiasi Pesawat Sinar-X Mammografi Di RSUD Kota Makasar, Skripsi, Jurusan Fisika, Fakultas Sains dan Teknologi Universitas Islam Negeri Alauddin Makasar, 2017.
Published
2021-10-17
How to Cite
YULIANTI, Ni Putu Pande Anita; SUTAPA, Gusti Ngurah; ASTINA, Kadek Yuda. Uji Kesesuaian Kualitas Berkas Sinar-X Menggunakan Half Value Layer (HVL) Filter Aluminium Pada Pesawat Sinar-X Stationary di Akademi Teknik Radiodiagnostik dan Radioterapi Bali. BULETIN FISIKA, [S.l.], v. 24, n. 1, p. 27–31, oct. 2021. ISSN 2580-9733. Available at: <https://ojs.unud.ac.id/index.php/buletinfisika/article/view/76312>. Date accessed: 21 nov. 2024. doi: https://doi.org/10.24843/BF.2023.v24.i01.p04.