QUALITY CONTROL OF GC PERFORMANCE FOR STEROID ANALYSIS
Abstract
This paper reports the process conducted to maintain a high quality of GC responses in obtaining reliabledata on steroid analyses of environmental samples. Works including setting up the Limit of Detection (LOD) and
Limit of Quantitation (LOQ), maintaining linearity, controlling the reproducibility, and determining its reliability are
described. It is evident that in this work the LOD of the GC is 0.0002 ng (by 1 mL injection of 0.2 mgL-1 samples) and
the LOQ is 0.001 ng (by 1 mL injection of 1 mgL-1 samples), which means it is capable of detecting much lower
concentrations of steroid compounds if the samples undergo 2000 times pre-concentration during preparation. Under
the range investigated the GC provides reliable quantitative linear responses to all steroid compounds.