PERBEDAAN KUALITAS FRESH FROZEN PLASMA YANG DICAIRKAN DENGAN METODE KONVENSIONAL DAN DENGAN METODE ALAT FFP THAWER

  • Didin Retno
  • Ahmad Riza Zainal
  • Budi D Machsoos
  • Djoko Heri Hermanto
  • Shinta Oktya Wardhani

Abstract

Saiful Anwar Hospital (SAH) doesn?t have standardized Fresh Frozen Plasma (FFP) liquefying tools. Standard Operating
Procedure (SOP) to liquify FFP at SAH is put the FFP on the table until diluted. Recently a new Thawer of FFP was
developed in SAH. In this study we compared between the two methods which are convensional and the new one. Thirtty
two bag of FFP were divided in two different treatments which is group of new Thawer and the convensional. Each
liqueed bag of FFP was recorded the time of liquefaction, aPTT, PPT and INR. Statistical analyses using independent T
test or Mann-Whitney as alternative. In this study found mean time of liquefaction of FFP on Thawer group were shorter
than conventional method 25.32 ± 3.4 vs 50.38 ± 5.52 minutes and statistically signicant (p < 0.05). Mean of aPTT on
new Thawer group is 36.79 ± 5.94; but on conventional method is 36.94 ± 5.53 second (p = 0.67). Mean of PPT 11.71 ±
0.95 second on new Thawer group and 11.84 ± 2.22 second on conventional method (p = 0.249). Mean INR each group
were 1.02 ± 0.08 on new Thawer group and 1.03 ± 0.19 on conventional method (p = 0.234). This study shows that new
FFP Thawer liquifay FFP more quickly than conventional method without decreasing the quality of FFP.

Downloads

Download data is not yet available.

Author Biographies

Didin Retno
Bagian/SMF Ilmu Penyakit Dalam FK Universitas Brawijaya/RS Saiful Anwar, Malang
Ahmad Riza Zainal
Bagian/SMF Ilmu Penyakit Dalam FK Universitas Brawijaya/RS Saiful Anwar, Malang
Budi D Machsoos
Bagian/SMF Ilmu Penyakit Dalam FK Universitas Brawijaya/RS Saiful Anwar, Malang
Djoko Heri Hermanto
Bagian/SMF Ilmu Penyakit Dalam FK Universitas Brawijaya/RS Saiful Anwar, Malang
Shinta Oktya Wardhani
Bagian/SMF Ilmu Penyakit Dalam FK Universitas Brawijaya/RS Saiful Anwar, Malang
How to Cite
RETNO, Didin et al. PERBEDAAN KUALITAS FRESH FROZEN PLASMA YANG DICAIRKAN DENGAN METODE KONVENSIONAL DAN DENGAN METODE ALAT FFP THAWER. journal of internal medicine, [S.l.], apr. 2014. Available at: <https://ojs.unud.ac.id/index.php/jim/article/view/3973>. Date accessed: 21 nov. 2024.
Section
Articles

Keywords

FFP, Thawer, conventional