Analisis ketidakpastian pengukuran pada mistar ingsut rentang 0-30 cm
Abstract
In science of measurement, an error is defined as difference result of measurement and the real value of the measurand.Actually, this real value is never precisely identified; consequently, the error is unknowns as well. In this case, the error canonly be estimated without knowing its real quantity. If a value will be considered as an estimated error, this value should be taken as an uncertainty. Uncertainty of Vernier Calliper determined with Type A and Type B evaluation, then the combinedstandard uncertainty is determined. Vernier Calliper that calibrated with Gauge Block class 1, the maximum uncertainty is ± 3.27 µm in dimension of 12.7 mm, with 95% confidence level.Downloads
Download data is not yet available.
How to Cite
AGUS SURYAWAN, I Gede Putu; KOMANG DWIJANA, I Gusti.
Analisis ketidakpastian pengukuran pada mistar ingsut rentang 0-30 cm.
Jurnal Energi Dan Manufaktur, [S.l.], nov. 2012.
ISSN 2541-5328.
Available at: <https://ojs.unud.ac.id/index.php/jem/article/view/2211>. Date accessed: 15 oct. 2025.
Issue
Section
Articles
Keywords
uncertainty in Measurement, Vernier Calliper